AGFA CR 30-X Service Manual page 256

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
Distance exaggerated
IMPORTANT:
When inserting the drawer unit adjustment tool, be sure, that the notch of the tool is
properly inserted in the right hand side of the cassette unit entry. See Figure 7.
In case the lateral stop is aligned with the plastic finger at the drawer unit there is no
Evaluation
adjustment required. Continue with section 2.1.3.
Otherwise perform lateral adjustment as described in next steps.
LATERAL ADJUSTMENT
(1)
Open the lateral position fixation
screws. See arrows.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
NOT OK
Drawer adjustment tool
517536ah.cdr
Figure 7
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Adjustments and Calibrations
OK
Lateral stop
Notch
Distance exaggerated
Figure 8
Agfa Company Confidential
NOT OK
517536dd.cdr
Chapter 3.6 / 8