AGFA CR 30-X Service Manual page 221

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(6)
Open 3 lower screws at front cover a few
turns.
(7)
Press two handles and pull out front cover
approx. 5 cm.
(8)
Disconnect cable at user interface board.
See Figure 196.
(9)
Disconnect ground cable at left hand side.
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
5175R3-5_004.cdr
Figure 195
Figure 196
Figure 197
Figure 198
Chapter 3.5 / 79
Agfa Company Confidential