DD+DIS150.06E
(5)
Disconnect the cable.
(6)
Remove the roller IP sensor.
ASSEMBLY:
(1)
Mount in reverse order.
(1)
Switch on the digitizer.
Verification:
(2)
Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3)
Insert cassette with unexposed IP in the digitizer.
(4)
At the workstation select examination type "System diagnosis – Flatfield".
(5)
Select "ID" to identify the image: Scanning starts.
(6)
Confirm that the image arrives at the workstation.
(7)
Reject the image.
The roller IP sensor is replaced.
Result:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 156
Chapter 3.5 / 65
Agfa Company Confidential