Replacing The Distributor Board - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
4.2

Replacing the Distributor Board

REQUIRED SPARE PART*:
CM+9 5175 1560 0
* The last digit in the spare part number indicates the spare
part revision at release of this document. When ordering,
the actual revision of the spare part is delivered.
REQUIRED TOOLS:
REQUIRED TIME:
30 min.
DISASSEMBLY:
(1)
Switch off the digitizer.
(2)
Disconnect the mains cable.
(3)
Simultaneously press the two buttons (A)
positioned underneath the feed table.
(4)
Pull out cassette unit (B).
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
3 mm
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Replacements / Repair Procedures
Figure 234
B
A
Figure 235
Agfa Company Confidential
A
Chapter 3.5 / 97