DD+DIS150.06E
(7)
Press both handles and pull out front cover
approx. 5 cm.
(8)
Disconnect cable at user interface board.
See Figure 105.
(9)
Disconnect ground cable at left hand side.
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
(12) Remove screw at erasure unit.
(13) Disconnect two cables at erasure unit.
(14) Take out erasure unit.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 105
Figure 106
Figure 107
Figure 108
Chapter 3.5 / 45