DD+DIS064.02E
Edition 1, Revision 4
Controls, Connections and Setup Procedures
Chapter
1
List of Contents
1
Safety .....................................................................1
1.1 General Safety Instructions ............................................... 1
1.2 Safety Instructions for Laser Products............................. 2
1.3 Safety Instructions for Cassette Unit ................................ 2
1.4 Safety Regulations ............................................................. 3
2
Preparations before Installation...........................5
2.1 Inspection of Packing......................................................... 5
2.2 Unpacking the Components .............................................. 6
2.3 Checking the Accessory Kit .............................................. 6
2.4 Removing the Shipping Brackets...................................... 7
2.5 Installation of Base Plate ................................................. 14
3
Controls and Connectors ...................................17
4
Operation Terminal Digitizer ..............................18
5
Requirements for Installation.............................19
5.1 General .............................................................................. 19
5.2 Required Service Programs............................................. 19
5.3 Preparations for the Installation...................................... 20
6
6.1 Electrical Connection ....................................................... 21
6.2 Start-up Procedure ........................................................... 22
6.3 Backup of the machine specific Data.............................. 24
7
Concluding the Installation ................................25
7.2 Last Steps.......................................................................... 28
8
9
ADC Compact Plus
(Type 5146)
Chapter 1 / I

Advertisement

Table of Contents
loading

Summary of Contents for AGFA ADC Compact Plus

  • Page 1: Table Of Contents

    5.1 General ................19 5.2 Required Service Programs..........19 5.3 Preparations for the Installation........20 Installation of the ADC Compact Plus Digitizer 21 6.1 Electrical Connection ............21 6.2 Start-up Procedure ............22 6.3 Backup of the machine specific Data......24 Concluding the Installation ........25...
  • Page 3: Safety

    · As is the case for all technical devices, the ADC Compact Plus must be operated, cared for and serviced correctly. · If you don't operate the ADC Compact Plus correctly or if you don't have it serviced correctly, Agfa-Gevaert is not liable for resulting disturbances, damages or injuries.
  • Page 4: Safety Instructions For Laser Products

    ADC Compact Plus. Safety Instructions for Laser Products The ADC Compact Plus is a class 1 Laser Product. It uses internally a 2 x 50mW laser diode, classification class IIIb. Under normal operating conditions - when both doors are closed - there can be no laser radiation outside the ADC Compact Plus.
  • Page 5: Safety Regulations

    A ground fault circuit interrupter (GFCI) is recommended, but not compulsory. Mains Connection Connection via plug and socket. Certificates · The ADC COMPACT Plus Digitizer is in compliance with EG regulation 93/42/EEC (Medical device). · The ADC COMPACT Plus Digitizer is tested in compliance with: ·...
  • Page 6 Controls, Connections and Setup Procedures DD+DIS064.02E Radio Interference Suppression It is hereby certified that the ADC COMPACT Plus Digitizer has interference suppression according to EN 55011: 1998 Class B as well as FCC Rules CR47 Part 15 Class A (North-America).
  • Page 7: Preparations Before Installation

    - dented edges - damage on the box - torn fixing elements (metal straps, screws). In case of damage visible from the outside turn to your local AGFA representative. · Checking of the attached safety indicators on the packing boxes.
  • Page 8: Unpacking The Components

    The individual boxes inside must not be opened by the forwarder. · Taking out the accessories or boxes and storing them in a safe place. · The ADC Compact Plus must be taken off the pallet and transported to the installation site.
  • Page 9: Removing The Shipping Brackets

    A: Unscrew the locking screws at the front of the digitizer. (8mm Allen key, extension included in accessory kit) · Then open the doors of the digitizer. Edition 1, Revision 4 ADC Compact Plus Chapter 1 / 7 (Type 5146)
  • Page 10 Unscrew the cassette unit from the frame (2 screws, 8mm allen key) Turn the cassette units through 90° Remove the two metal brackets (6 screws on the front and rear side of the cassette units, 7mm socket wrench) Chapter 1 / 8 ADC Compact Plus Edition 1, Revision 4 (Type 5146)
  • Page 11 B: Cut the cable tie on the deflector of the output buffer. C: Cut the cable tie on the feed gate of the input buffer. D: Screw on the two adjustable bases (7mm socket wrench). Edition 1, Revision 4 ADC Compact Plus Chapter 1 / 9 (Type 5146)
  • Page 12 Controls, Connections and Setup Procedures DD+DIS064.02E Removing shipping brackets of transport units 514601EE.CDR Figure 4 · Unscrew two metal plates (6 screws, 8 mm Allen key) Chapter 1 / 10 ADC Compact Plus Edition 1, Revision 4 (Type 5146)
  • Page 13 Controls, Connections and Setup Procedures DD+DIS064.02E 514601ej.cdr A: Remove the cushioning block between IP –transport unit postscan and scan unit (hardly visible). B: Cut and remove the cable tie. Edition 1, Revision 4 ADC Compact Plus Chapter 1 / 11 (Type 5146)
  • Page 14 A: Slide out the IP – transport unit left (postscan), cut the two cable ties and remove the plastic plate. - Slide out the IP – transport unit right (prescan) B: - Remove the polystyrene blocks from both transport units very cautiously. Chapter 1 / 12 ADC Compact Plus Edition 1, Revision 4 (Type 5146)
  • Page 15 B: Mount the light tight tin plate (included in accessory kit) by fastening six Phillips screws. Remove the polystyrene blocks jammed at the feed- and exit rollers of the scan unit (2 pieces). Edition 1, Revision 4 ADC Compact Plus Chapter 1 / 13 (Type 5146)
  • Page 16: Installation Of Base Plate

    A: Mount the base plate with four screws on the floor (recommended, but not mandatory), see arrows. Figure 8 B: Move the digitizer onto the base plate. Take care that the rollers are guided by the rail! Chapter 1 / 14 ADC Compact Plus Edition 1, Revision 4 (Type 5146)
  • Page 17 D: Turn screws (on both sides, left and right) by hand to their stop, then screw it in one more turn. Do not give more turns! Figure 10 E: Put the plastic caps on the screws to protect them. Edition 1, Revision 4 ADC Compact Plus Chapter 1 / 15 (Type 5146)
  • Page 18 Controls, Connections and Setup Procedures DD+DIS064.02E Figure 11 F: Adjust the two bases (7 mm socket wrench) Chapter 1 / 16 ADC Compact Plus Edition 1, Revision 4 (Type 5146)
  • Page 19: Controls And Connectors

    3. Port for the Service PC (RS 232 - 25-pin D connector, female D) 4. Network connection (Twisted Pair) 5. Mains cable VDE/UL 6. Grounding cable connection Edition 1, Revision 4 ADC Compact Plus Chapter 1 / 17 (Type 5146)
  • Page 20: Operation Terminal Digitizer

    (from right digit to left order (from left digit to right digit) and toggles between digit) and toggles between values in a field values within a field. Chapter 1 / 18 ADC Compact Plus Edition 1, Revision 4 (Type 5146)
  • Page 21: Requirements For Installation

    The latest version of the CCM - Tool can be downloaded from MedNet. The CCM - Tool is needed to create and modify the configuration file (adc.cpf) for the ADC Compact Plus and the other ADC System Components. If there is already a previous version installed on your Service PC, please remove the complete directory as well as the icons.
  • Page 22: Preparations For The Installation

    We recommend, to create and adapt the configuration file adc.cpf, which is used by all ADC System Components, before starting the installation. This allows you a smooth installation. For detailed information see ADC System Components, DD+DIS198.00E, chapter 6.2. Chapter 1 / 20 ADC Compact Plus Edition 1, Revision 4 (Type 5146)
  • Page 23: Installation Of The Adc Compact Plus Digitizer

    Installation of the ADC Compact Plus Digitizer Electrical Connection 6.1.1 Connecting the Power Cord ADC Compact Plus comes with a UL and a VDE power cord. · Connect the power cord to the machine and to the power supply. 6.1.2 Voltage Adjustment ADC Compact Plus has an automatic voltage selector.
  • Page 24: Start-Up Procedure

    Controls, Connections and Setup Procedures DD+DIS064.02E Start-up Procedure 6.2.1 Mechanical Start-up ADC Compact Plus must not be connected to a network during the start-up process. Mechanical Start-Up Start-Up: 1. Switch on the digitizer. 2. Check the operation terminal during the initialization.
  • Page 25 DD+DIS064.02E Controls, Connections and Setup Procedures 6.2.2 Integrating ADC Compact Plus into an independent, pure ADC Network Pure ADC Network (minimum configuration) · 1 ADC Compact Plus digitizer · 1 Processing Station · 1 Preview/ID Station In this case the ex factory settings for the ip- address (adcc1 192.9.200.199) are being used.
  • Page 26: Backup Of The Machine Specific Data

    <Machine specific data> 4. Start the backup procedure. 5. Label the floppy as follows: S/N: <XXX> Date: <date> Software- ACP_xxxx version: End of Backup (approx. 15 min). Chapter 1 / 24 ADC Compact Plus Edition 1, Revision 4 (Type 5146)
  • Page 27: Concluding The Installation

    For exposure place the cassette in length direction to the X-ray tube see figure above. - 1st exposure: Doses: 20 mAs-77kV-1.3 m distance,1.5 mm Cu filter - Turn cassette through 180°. - 2nd exposure: parameters as in 1st exposure. Edition 1, Revision 4 ADC Compact Plus Chapter 1 / 25 (Type 5146)
  • Page 28 · Print the testsheets via Processing Station (window 0.6). · Insert a test sheet (CM+7.8966.1235.0) into the cassette and expose with the following parameters: 10 mAs - 77 kV - approx. 1,3m distance . Chapter 1 / 26 ADC Compact Plus Edition 1, Revision 4 (Type 5146)
  • Page 29 After having finished the test, it is strongly recommended to delete the test sheets on the hard disk of the Processing Station. · Proceed according to the subsequent evaluation criteria: Edition 1, Revision 4 ADC Compact Plus Chapter 1 / 27 (Type 5146)
  • Page 30: Last Steps

    Installation of other ADC System Components For detailed information on the installation of other ADC System Components like Processing Station or ID –Station, see: ADC System Components, DD+DIS198.00E, Chapter 3. Chapter 1 / 28 ADC Compact Plus Edition 1, Revision 4 (Type 5146)
  • Page 31: Checklist For Installation / Machine Start -Up

    Set in Maintenance Menu: network settings Connect Processing Station to the network Check basic functions like receiving, printing sending, viewing images Reset Infocounters Create a site specific backup Edition 1, Revision 4 ADC Compact Plus Chapter 1 / 29 (Type 5146)
  • Page 32 Fill in the form "Site and System Data" and send it to MICSO Send the completed "Installation Reports" to the regional CSO manager Remote Access Make connection to Service Host Chapter 1 / 30 ADC Compact Plus Edition 1, Revision 4 (Type 5146)