AGFA CR 30-X Service Manual page 418

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS172.06E
WARNING:
Using a FWI cable > 4.5 m may lead to unstable or no communication
between digitizer and NX workstation.
Only use the FireWire cable which is part of delivery. Longer cables are not
supported.
WARNING:
When different combinations of equipment are used in various medical
environments a potential difference (V) can exist between the protective
earths in different localities. If the protective earthing fails this potential
difference can cause a HAZARD for the OPERATOR or for the PATIENT.
To comply with ISO 60601-1 (annex I) all computers and peripherals must be
connected to the same power source.
Always connect the associated monitor to the same Uninterruptible Power
Source as the PC.
When connecting digitizer and control PC at two different sockets, installed in
the wall: Inform the hospital technician that digitizer and control PC have to be
connected to the same ground and the same phase.
WARNING:
Excessive vibrations during scanning may decrease image quality.
The structure and stability of the used table need to be suitable in relation with the
size and weight of the system. The table may not be subject to excessive shock
and vibrations from other sources.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2010
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Installation Planning
Chapter 11 / 16
Agfa Company Confidential
1