Page 1
HealthCare Maintenance Order-No.: DD+DIS014.02E *1UT4741* ADC Compact Type 5145 / 100/101/140 1 Piece UT474 MA1 Type 5145 / 200/201/240 Edition ADC Compact Digitizer Type 5145 / 100/101/140 Type 5145 / 200/201/240 Base: VME This documentation is separately available. Order No: DD+DIS014.02E...
Page 2
The CE Declaration (CE Conformity) becomes invalid if the product is changed without explicit consent of the manufacturer! This applies to all parts, not only to safety elements. We reserve the right to technical changes Section 12 ADC Compact Edition 2, Revision 8 (Type 5145 / 100/101/140/200/201/240)
Checking the Image Quality........12 Test Cycles................12 Exposure of a Flatfield ............12 Evaluation of a Flatfield ............15 Completion of Maintenance ........16 Maintenance Checklist Edition 2, Revision 8 ADC Compact Section 12 / I (Type 5145 / 100/101/140/200/201/240)
· If you don't operate the ADC Compact correctly or if you don't have it serviced correctly, Agfa-Gevaert is not liable for resulting disturbances, damages or injuries. · When installing the ADC Compact, care must be taken to ensure that there is either a mains plug or an all-cable disconnecting device in the internal installation fitted near the ADC Compact and that it is easily accessible.
"must" maintenance points. Maintenance Frequency The maintenance has to be carried out: · every 15.000 cycles or · every 6 month Required Time approximately 3 h Section 12 / 2 ADC Compact Edition 2, Revision 8 (Type 5145 / 100/101/140/200/201/240)
2.2 Throughput For throughput most important are the cycles per day. They usually count between 50 and 200. Section 12 / 4 ADC Compact Edition 2, Revision 8 (Type 5145 / 100/101/140/200/201/240)
(1) Open the doors of the digitizer while the machine is switched (2) As soon as the left door is opened only a little all assemblies must be deenergized. 514512AL.CDR Figure 3 Section 12 / 6 ADC Compact Edition 2, Revision 8 (Type 5145 / 100/101/140/200/201/240)
(4) Dust the VME Rack itself, the housing of the rack and all fan-grids. Handle the cables with care! Danger of damage by unplugging and Figure 6 plugging. Edition 2, Revision 8 ADC Compact Section 12 / 7 (Type 5145 / 100/101/140/200/201/240)
(3) Exchange suction cups once a year. (4) In case of vacuum problems, check the suction cups by bending up the edges (check for tears) and replace if necessary. Figure 8 Section 12 / 8 ADC Compact Edition 2, Revision 8 (Type 5145 / 100/101/140/200/201/240)
2 x 1000 W lamps (CM+9.0450.6558.0) · 1 x 2000 W lamps (CM+9.0450.6559.0) (1) Remove the complete Erasure Unit by following the sequence and arrows in figure beside. Figure 11 Edition 2, Revision 8 ADC Compact Section 12 / 9 (Type 5145 / 100/101/140/200/201/240)
Page 13
Do not touch the lamps with your bare fingers! Use a soft cloth to insert the lamp. 514512cx.cdr Figure 12 (7) Re-insert the Erasure unit. Section 12 / 10 ADC Compact Edition 2, Revision 8 (Type 5145 / 100/101/140/200/201/240)
Visual Check (1) Check if there are scratches on the surface. (2) Check if edges are loose as an indication for mechanical problems at IP transport. Edition 2, Revision 8 ADC Compact Section 12 / 11 (Type 5145 / 100/101/140/200/201/240)
<Send flatfield> <Calibration pattern> <Banding pattern> Print the flat fields "Calibration" and "Banding" via the Processing Station (window setting of 0.6, without changing the level setting). Section 12 / 12 ADC Compact Edition 2, Revision 8 (Type 5145 / 100/101/140/200/201/240)
Page 16
In the <Sub-examination> list, click <Flat field>. · Confirm the Exposure class <200>. Make sure that the outlined areas are filled in as shown in the example. Edition 2, Revision 8 ADC Compact Section 12 / 13 (Type 5145 / 100/101/140/200/201/240)
Page 17
(4) Insert the cassette into the digitizer and print the image on a printer with a window setting of 0.6 without changing the level setting. Figure 14 Section 12 / 14 ADC Compact Edition 2, Revision 8 (Type 5145 / 100/101/140/200/201/240)
Fine sharp white or gray lines in fast scan direction on the flat field (see beside). · Check diagnostic images. If there are still unacceptable effects please contact the Support Center. Figure 16 Edition 2, Revision 8 ADC Compact Section 12 / 15 (Type 5145 / 100/101/140/200/201/240)
(2) Make a backup of the system on floppy (3) Inform the customer about what was done during the maintenance and which repairs need to be done in next future. Section 12 / 16 ADC Compact Edition 2, Revision 8 (Type 5145 / 100/101/140/200/201/240)
Page 20
AGFA and the Agfa-Rhombus are trademarks of Agfa-Gevaert AG, Germany Herausgegeben von / Published by / Edité par / Pubblicato da / ../ Editado por Agfa-Gevaert AG Fototechnik Tegernseer Landstraße 161 D - 81539 München Printed in Germany / Imprimé en Allemagne /...
Page 21
HealthCare Maintenance Checklist Order-No.: DD+DIS014.02E ADC Compact Type 5145 / 100/101/140 Type 5145 / 200/201/240 Work Instruction for order no. Cycles The maintenance has to be carried out: · Every 6 months or every 15 000 cycles Maintenance must be carried out according to the maintenance instructions, see section 12 of your Service Documentation.
Done General · Confirm the maintenance · Make a backup of the system on floppy · Inform the customer about maintenance and which repairs are necessary Maintenence Checklist ADC Compact Edition 2, Revision 8 Page 2 (Type 5145 / 100/101/140/200/201/240)