Replacing The Roller Stepper Motor (M5) - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(1)
Insert cassette with unexposed IP in the digitizer.
Verification
(2)
At the workstation select examination type "System diagnosis – Flatfield".
(3)
Select "ID" to identify the image: Scanning starts.
(4)
Confirm that the image arrives at the workstation.
(5)
Reject the image.
2.18

Replacing the Roller Stepper Motor (M5)

REQUIRED SPARE PART*:
CM+9 5175 6350 0
* The last digit in the spare part number indicates the spare
part revision at release of this document. When ordering,
the actual revision of the spare part is delivered.
REQUIRED TOOLS:
REQUIRED TIME:
30 min.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
3 mm,
with ball-shaped head
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Replacements / Repair Procedures
Figure 131
Agfa Company Confidential
Chapter 3.5 / 55