Replacing The Pmi Board - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(7)
Disconnect all cables at the PMI Power
distribution board.
(8)
Open the 6 mounting screws.
(9)
Remove the PMI Power distribution board.
ASSEMBLY:
(1)
Mount in reverse order.
(1) Switch on the digitizer.
Verification:
(2) Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3) Insert cassette with unexposed IP in the digitizer.
(4) At the workstation select examination type "System diagnosis – Flatfield".
(5) Select "ID" to identify the image: Scanning starts.
(6) Confirm that the image arrives at the workstation.
(7) Reject the image.
The PMI Power Distribution Board is replaced.
Result:
3.4

Replacing the PMI Board

The replacement instructions for the PMI Board, spare part number*
CM+9 5175 2180 0, are enclosed to the spare part.
* The last digit in the spare part number indicates the spare part revision at release of this document.
When ordering, the actual revision of the spare part is delivered.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
5175R3-3_103
Figure 219
Chapter 3.5 / 89
Agfa Company Confidential