DD+DIS150.06E
6.1.4

Scanning the Flatfield Image

(1)
At the NX Workstation select examination type "System diagnosis – Flatfield".
(2)
Insert exposed cassette in the digitizer.
(3)
Scan cassette and wait until the digitizer shows the ready screen.
(4)
Print the image(s) or send it to the archive (depending on what is available).
6.1.5

Checking the Image at the Lightbox or Viewing Station

NOTE:
Checking the image may only happen on lightboxes and viewing stations that comply
with the specified installation and configuration conditions for diagnostic reading. For
details refer to the respective instructions of the viewing stations / light boxes.
Check the image quality of the flatfield for stripes and large area inhomogeneities.
Result:
If the scanned flatfield does not show stripes in fast scan or slow scan direction or
large area inhomogeneities no further action is required.
If the flatfield shows stripes in fast scan or slow scan direction or large area
inhomogeneities refer to the Troubleshooting Guide, DD+DIS151.08E. See also
NOTE below.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2009
Controls, Connections, and Setup Procedures
NOTE:
For external partners accessing the "Troubleshooting Guide" on
MedNet, GSO Library may be restricted. In such a case please contact
your local Agfa representative.