NOTE:
The laser beam switches off automatically after 30 minutes.
(16) Check the LEDs on the distribution board.
(17) Evaluate as follows (see table below):
Evaluation of the LEDs on the distribution board:
#
Status
1
LED 1 ON
2
LED 1 ON
3
LED 1 ON
4
LED 1 ON
5
LED 1
OFF
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
LED 2 ON
LED 3 ON
LED 2 OFF
LED 3 ON
LED 2 ON
LED 3 OFF
LED 2 OFF
LED 3 OFF
LED 2 OFF
LED 3 OFF
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Adjustments and Calibrations
LED 3
Switch
Figure 33
Action
No further action required. Laser beam
position OK. Continue with next step.
Laser beam position adjustment
required. Continue with section 2.5.2.
Check, whether laser beam is really
switched on: Look from the digitizer
right hand side, whether the red laser
line is visible. For more information
refer to the troubleshooting guide,
DD+DIS151.08E.
If laser beam is visible, the laser beam
position adjustment is required.
Continue with section 2.5.2.