Replacing The Drawer Unit; Replacing The Erasure Unit - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
Verification:
(1)
Switch on the digitizer.
(2)
Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3)
Insert cassette with unexposed IP in the digitizer.
(4)
At the workstation select examination type "System diagnosis – Flatfield".
(5)
Select "ID" to identify the image: Scanning starts.
(6)
Confirm that the image arrives at the workstation.
(7)
Reject the image.
The Chip Reader is replaced.
Result:
2.8

Replacing the Drawer Unit

The replacement instructions for the Drawer Unit, spare part number*
CM+9 5175 6100 4, are enclosed to the spare part.
* The last digit in the spare part number indicates the spare part revision at release of this document.
When ordering, the actual revision of the spare part is delivered.
2.9

Replacing the Erasure Unit

REQUIRED SPARE PART*:
CM+9 5175 3600 1
* The last digit in the spare part number indicates the spare
part revision at release of this document. When ordering,
the actual revision of the spare part is delivered.
REQUIRED TOOLS:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
3 mm,
with ball-shaped head
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Replacements / Repair Procedures
Agfa Company Confidential
Figure 93
Chapter 3.5 / 38

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents