Performing Stall Calibration For All Motors; Check Of Safety Switch; Performing Test Cycles; Technical Image Quality Check Of The System - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS171.06E
Maintenance points - Power off
Visual Inspection
Cleaning; General
Cleaning of the
Erasure Unit
Checking the Teflon
tape at the cassette
unit
Greasing the spindle
nut
Cleaning of the Scan
Line
Cleaning of scan
rollers
Checking of Cassette
Condition
Maintenance points - Power on
Performing stall
calibration for all
motors.
Check of safety switch • Opening of cassette unit to check function of
Performing Test
Cycles
Technical Image
Quality Check of the
System
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
• Visual check of
o Power and FireWire cable
o Covers
o Cables inside the digitizer, especially at cassette
unit left hand side
o 4 belts in the cassette unit
• Cleaning the inside with a vacuum cleaner
Especially checking for loose dust in transport unit
• Cleaning of the erasure unit with a lint free cloth
• Checking the Teflon tape at the cassette unit and
replacement if required
• Put some grease at the spindle in the area of the
spindle nut.
• Cleaning of the Scan Line if dust stripes are visible
• Cleaning of the 2 lower scan rollers if spot artifacts
are visible on the image.
• Check of all cassettes for damage

• Performing stall calibration for all motors

safety switch
• Performing of 5 test cycles per cassette format
• Exposure and check of flatfield.
Comparison with flatfield made before
maintenance
• Comparison with test pattern "lowest acceptable
quality" (slow scan / fast scan direction) if image
quality is not OK.
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Maintenance
Maintenance Checklist
Chapter 9 / 32
Agfa Company Confidential
Done
Done