Error 12800 After Switch-On Of The Digitizer (Hq_0901140001); Additional, Minor Bug Fixes - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
Solution
If software CRE_1607A is installed:
In Software CRE_1607 the digitizer is automatically registered when it gets connected to
the control PC. It is not required anymore to register the digitizer in service client menu
"Configuration - Site specific Data" during installation.
In addition, the digitizer detects in most cases automatically, whether a board or a
module was exchanged: Error message 4642 ("A new module serial number has been
detected") will only appear in exceptional cases.
Solution
If software CRE_1607A is not installed yet:
If the problem appears at a digitizer with software < CRE_1607, a software upgrade in
the error situation will solve most problems if digitizer was registered during the
installation procedure: In software CRE_1607 an algorithm is implemented to repair
corrupted e-labels.
11.2.3

Error 12800 after Switch-on of the Digitizer (HQ_0901140001)

Symptom
Error 12800 ("Procedure selftest of Clamper terminated with failure") during or after boot
up of the Digitizer. The fault can be sporadic or persistent.
Cause
Control parameter of the roller clamper drive (motor M6) test does not match in some
cases with certain design parameter of the roller clamper drive.
Solution
Improved selftest routine during boot up in the software CRE_1607A
11.2.4

Additional, minor Bug Fixes

The following lists minor software issues, which have been solved with CRE_1607A:
The sporadic error 10752 at boot-up is solved by a time delay to test light
incidence.
The menu 'Diagnose Cycle' in the service client cannot be accessed anymore
remotely in software CRE_1607A.
No user message was displayed, if the FireWire link to the control PC was
interrupted during the start of scan and after return to the cassette.
Now software CRE_1607A shows a message and asks the user to check, whether
the image plate was proper scanned and erased.
Sporadic error 2595 (Image transmission failed. System is retrying. If image does
not appear, please restart the PC) despite of successful image transmission.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
01-2011
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Repair and Service
Software Releases and Patches
Agfa Company Confidential
Chapter 3.8 / 21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents