DD+DIS150.06E
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
ASSEMBLY:
(1) Mount in reverse order.
(1) Switch on the digitizer.
Verification:
(2) Wait till the green light at the user interface lights up
(takes approx. 1,5 min.).
The front cover is replaced.
Result:
1.2
Replacing the Left Cover
REQUIRED SPARE PART*:
CM+9 5175 1503 1
* The last digit in the spare part number indicates the spare
part revision at release of this document. When ordering,
the actual revision of the spare part is delivered.
REQUIRED TOOLS:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
3 mm,
with ball-shaped head
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Replacements / Repair Procedures
Agfa Company Confidential
Figure 7
Figure 8
Chapter 3.5 / 7