Service Client Menu "Exchanged Components - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
2.2.16

Service Client Menu "Exchanged Components"

Menu "Exchanged components" allows to define whether the exchanged component
Purpose
is a module (e.g. slow scan unit) or part of a module (e.g. slow scan board).
Only available as of software CRE_1509.
(1) Select "Configuration – Exchanged Components".
Usage
(2) Select "Board" or "Module" according to following table:
Exchanged component
Slow scan unit
Slow scan board
PMT with light collector
IP handling board
Cassette fixation unit
Optic module
PMI2 / PMI 3 board
NOTE:
This menu is required, as for some replacements it is not known for the digitizer,
whether the exchanged component is a module (e.g. slow scan unit) or part of a
module (e.g. slow scan board). For more details refer to sections 2.2.16.1,
Replacement Scenarios and 2.2.16.2, Handling of Replacements in different Software
Versions
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
Type of Exchange
Figure 23: Example: Selecting Board as "Type of Exchange"
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Tools and auxiliary Means
Module
Board
Module
Board
Module
Module
Board
Agfa Company Confidential
Chapter 3.2 / 28