DD+DIS172.06E
Installation
guidelines
Mains connection
in the installation
room
Mains cable
(part of delivery)
IMPORTANT:
The digitizer is delivered with two power cables for US/Europe. If you need a different
cable, organize it locally.
Fire Wire cable
Mains switch
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2010
Electrical installations have to be done in accordance with
national regulations or statutory requirements, e.g.:
(Germany) Electrical installations in the installation
VDE:
room must be in compliance with the regulations IEC
364, VDE 0100, and VDE 0107.
(US) Electrical installations in the installation room
UL:
must be in compliance with the regulations: "National
Electrical Code" (NEC) (NFPA70).
Double earthed pin outlet in compliance with DIN
VDE:
49441 and with CEE 7 standard cover V II.
Earthed contact outlet for the NEMA 5-15 P plug
UL:
CEE7/7 - H05VV -
Euro-
Version :
F3G1 - 3 x 1 mm² with
IEC C13 plug (female);
cable length 3 m
NEMA 5-15P - SJT,
US/CAN-
Version :
3 x AWG 18 with
IEC C13 plug (female);
Cable length 3 m,
Hospital grade (HG)
IEEE 1394, - 4,5 m with ferrite core (part of digitizer delivery)
Upon digitizer installation, it must be ensured that either the
mains connector or an all-pole circuit breaker for the installation
on site is located close to the machine and easily accessible.
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Installation Planning
IEC C13
NEMA 5-15P
Chapter 11 / 19
Agfa Company Confidential
1
2
3
CEE7/7
5175_C11_001.cdr
IEC C13