DD+DIS150.06E
(8)
Pull at the toothed belt to open
the cassette opener.
CAUTION:
Misalignment of drawer unit possible when pushing it at the upper plate.
When pushing the drawer unit only do this via drawer unit axle.
(9)
Push drawer unit at the axle
completely to the front.
2.1.2
Checking and Adjusting the lateral Position
(1)
Insert the drawer unit adjustment
tool in the cassette unit entry.
(2)
Push it completely to the right.
(3)
Check, whether plastic finger at
drawer unit is aligned with the
lateral stop of the drawer
adjustment tool. See Figure 7.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
Adjustments and Calibrations
Upper plate
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
517536ak.cdr
Figure 4
Drawer axle
Figure 5
Figure 6
Chapter 3.6 / 7
Agfa Company Confidential