Connecting The Cables - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(1)
Remove two screws in the rear.
(2)
Mount distance holder.
4.2

Connecting the Cables

WARNING:
Using a FWI cable > 4.5 m (15 ft) may lead to unstable or no communication
between digitizer and NX workstation: Retakes possible.
To connect digitizer and control PC, only use a standard 6-pin FWI cable with a max.
length of 4,5 m or 15 ft respectively.
WARNING:
Using different power circles may lead to ground potential differences: In this
case the leakage current via FireWire cable may exceed the limits as defined
by IEC 60601:
Confirm that both, control PC and digitizer are connected to the same ground, e.g.
via multiple socket or UPS.
When connecting digitizer and control PC at two different sockets, installed in the
wall: Inform the hospital technician that digitizer and control PC have to be
connected to the same ground and the same phase.