Service Client Menu "Diagnose Cycle - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
2.2.8

Service Client Menu "Diagnose Cycle"

Menu "Diagnose Cycle" allows to perform a complete scan cycle manually step by
Purpose
step e.g. for troubleshooting purpose.
A) Diagnose Cycle of the complete device:
Usage
(1) Select "Analysis & Repair - Diagnose Cycle".
(2) Put a cassette into the digitizer.
(3) Activate one of the grey displayed actions.
B) Diagnose Cycle of the Cassette Unit only:
(1) Open the cassette unit.
(2) Insert a service key in the safety switch.
(3) Select "Analysis & Repair - Diagnose Cycle".
(4) Put a cassette into the digitizer.
(5) Activate one of the grey displayed actions.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
Figure 15
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Tools and auxiliary Means
Agfa Company Confidential
Chapter 3.2 / 20