7.2

Transport and Storage Conditions

Transport of packed
device:
Ambient temperature
during transport:
Storage of packed device:
Ambient temperature
during storage
Relative humidity during
storage:
Rate of temperature
change during storage:
8

Electrical Connections

Warning:
Electrical device. Shock possible.
INSTRUCTION:
Do not position Agfa EQUIPMENT so it is difficult to operate the disconnection
device when an APPLIANCE COUPLER or separable plug is used as
isolation.
Local and International wiring regulations must be observed. Check all
supplies and voltages, currents, trips and fuses with the Hospital facilities
department or their engineers.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2010
According IEC 721-3-1 Transport: 2K2, 2M2
Minimum:
Maximum:
According IEC 721-3-1 Storage: 1K2, 1M2
Minimum:
Maximum:
Minimum:
Maximum:
Maximum:
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Installation Planning
-25 °C (-13 °F) for maximal 72 h
+55 °C (131 °F) for maximal 96 h
+5 °C (41 °F)
+40 °C (104 °F)
5 %
85 %
0,5 °C/min
Agfa Company Confidential