(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
(12) Remove four screws.
(13) Remove the user interface board.
ASSEMBLY:
(1)
Mount in reverse order.
(1)
Switch on the digitizer.
Verification:
(2)
Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3)
Insert cassette with unexposed IP in the digitizer.
(4)
At the workstation select examination type "System diagnosis – Flatfield".
(5)
Select "ID" to identify the image: Scanning starts.
(6)
Confirm that the image arrives at the workstation.
(7)
Reject the image.
The user interface board is replaced.
Result:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures