DD+DIS150.06E
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
(15) Slide IP guide plate to the left and lift it out.
(16) Remove screws at magnets.
(17) Remove magnets.
ASSEMBLY:
(1)
Mount in reverse order.