DD+DIS172.06E
7
Climatic and ambient Conditions
7.1
Operating Conditions
Ambient temperature:
Relative humidity:
At 25°C (77°F) ambient
temperature
Magnetic fields:
Radiation influences:
Warning:
Image plate is sensitive for X-rays. Poor image quality possible.
The digitizer and the cassette storage shall be protected against X-ray radiation this
way, that the annual dose equivalent at the installation place will not exceed 1 mSv.
Ambient Light:
High Frequency Emission
and Immunity:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2010
Min.:
Max.:
Best:
Min.:
Max.:
Best:
Max. 3.8 µT in conformance with EN6100-4-8:
level 2
If the digitizer is installed in the X-Ray room, it
has to be protected against scatter radiation.
The digitizer may not be operated in direct sun
light.
The image quality will not be affected, if
operated in ambient light of max. 2500 Lux.
For detailed information refer to the digitizer
user manual , Appendix B.
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Installation Planning
+15 °C (59 °F)
+30 °C (86 °F)
+20 °C (68 °F) to 25° C (77°F)
15 %
75 %
30 % to 60 %
Chapter 11 / 17
Agfa Company Confidential
1
2