DD+DIS150.06E
4.3.2
Electrical Check according to National Regulations
(1) Perform the electrical check according to National Regulations.
Example for Germany: BGV A3 and DIN VDE 0751 or subsequent regulations.
4.3.3
Verification of the successful Power Supply Replacement
(1) Switch off/on the digitizer.
Verification:
(2) Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3) Insert cassette with unexposed IP in the digitizer.
(4) At the workstation select examination type "System diagnosis – Flatfield".
(5) Select "ID" to identify the image: Scanning starts.
(6) Confirm that the image arrives at the workstation.
(7) Reject the image.
The power supply is replaced.
Result:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Chapter 3.5 / 105
Agfa Company Confidential