4.3.2
Electrical Check according to National Regulations
(1) Perform the electrical check according to National Regulations.
Example for Germany: BGV A3 and DIN VDE 0751 or subsequent regulations.
4.3.3
Verification of the successful Power Supply Replacement
(1) Switch off/on the digitizer.
Verification:
(2) Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3) Insert cassette with unexposed IP in the digitizer.
(4) At the workstation select examination type "System diagnosis – Flatfield".
(5) Select "ID" to identify the image: Scanning starts.
(6) Confirm that the image arrives at the workstation.
(7) Reject the image.
The power supply is replaced.
Result: