AGFA CR 30-X Service Manual page 275

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(5)
Fasten the four optic module
mounting screws.
(6)
Once more check the red LED 2 and
LED 3: They should stay light up
when all four optic module mounting
screws are fastened.
CAUTION:
If switch on distributor board stays in "ON" position during digitizer operation,
reduced image quality possible.
Set the switch on the distributor board to the OFF position before handing the digitizer
over to the customer.
(7)
Set the switch on the distributor
board to OFF position: The yellow
LED 1 does not light up anymore.
(8)
Mount the top cover.
(9)
Close the cassette unit.
(10) Connect the cables in the rear:
Observe the FireWire connector
direction. See Figure 40.
CALIBRATION:
(1)
Perform a shading calibration. Refer to section 3.1 of this chapter.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 39
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:
CAUTION! Remove power plug before servicing.
(1) This device may not cuase harmful interference (2) This device must accept any interference
Transport may only be undertaken as described in the operator manual.
received, including interference that may cause undesired operation.
This Class A digital apparatus meets all requirements of the Canadian Interference Causing Equipment
ACHTUNG! Vor Öffnen des Gerätes Netzstecker ziehen.
Regulations.
Cet appareil numerique de la classe A respecte toutes les exigences du Reglement sur le material
Vor dem Transport unbedingt Hinweise im Bedienhandbuch beachten.
brouillant du Canada.
ATTENTION! Retirez la prise de courant avant l'entretien.
Type B equipment
Cat appareil être transporté confrom au manual d'operations.
Geräte Typ B
Appareil du type B
The manufacturer certifies that this
product conforms to all applicable
provisions of 21 CFR Subchapter J in
effect of the date of manufacture
Man ufa ctur er
AGFA HEALTHCARE N.V.
Sep testr aat 27 - 26 40 - Mor tsel - BE L GIUM
WI T H R E S P E C T T O
ME D I C A L E L E C T R I C A L
E QU I P ME N T
E L E C T R I C S H OC K , F I R E ,
A C C OR D A N C E WI T H
H A Z A R D S ON L Y I N
A N D ME C H A N I C A L
C
U
R
L
U S
Made in Germany
C A N / C S A C 22. 2 N O. 6 01. 1
U L 6 060 1-1 /
9F 03
P eis senberg
MMMMM-Y YY Y
FireWire
Connector
Direction
Figure 40
Chapter 3.6 / 27
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents