DD+DIS150.06E
(16) Slide IP guide plate to the left and lift it out.
(17) Remove cable at plug P29 of IP handling
board.
(18) Open cable holders.
(19) Remove 4 screws of stepper motor roller
clamping (only 3 screws are shown in
Figure 152).
(20) Remove motor.
ASSEMBLY:
(1)
Mount in reverse order. Observe IMPORTANT note below:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 150
Figure 151
Figure 152
Chapter 3.5 / 62