DD+DIS171.06E
3.6

Cleaning of the Scan Line

IMPORTANT:
Only clean the scan line, if vertical stripes are visible in the image made under step
2.6, Checking current Image Quality.
Otherwise skip this step.
REQUIRED TOOLS:
Cleaning brush (attached to the digitizer "Cassette Unit")
(1)
Clean the scan line by wiping with
the cleaning brush a few times over
the whole scan length.
The last movement must be
continuous from rear to front.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Maintenance
Figure 20