DD+DIS171.06E
2.6
Checking current Image Quality
REQUIRED TOOLS:
•
Lint free cloth
•
ADC screen cleaner
•
Cu filter
(1)
Select the largest available cassette.
If possible, use a 35x43 cm Genrad cassette.
If in size 35x43 cm FLFS (Full Leg / Full Spine) cassettes are available only, use
a FLFS cassette.
NOTE:
FLFS cassettes can be recognized by a
sticker on the cassettes.
(2)
At the digitizer select the "Erase key".
(3)
Insert the selected cassette: The
status LED is blue flashing.
(4)
When finished (status LED changes
to green), remove the cassette.
(5)
Open the cassette using a key or a
pen.
(6)
Check the image plate for dirt or
scratches.
If dirt is visible, clean the image plate
with ADC screen cleaner and a soft
lint-free cloth.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Maintenance
Genrad Cassette
FLFS
Figure 3
517509ag.cdr
Figure 4
Figure 5
Agfa Company Confidential
FLFS Cassette
FLFS
Chapter 9 / 11