Software Patch Cre_1309A; Solved Bugs With Software Cre_1309A Compared To Cre_1309 - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
6

Software Patch CRE_1309A

Type of new Software
Release Date
Digitizer SNs
Availability
Prerequisites
Reference Document
*Note:
On MedNet, GSO Library always the latest available software release can be
downloaded.
6.1

Solved Bugs with Software CRE_1309A compared to CRE_1309

Solved sporadic error codes 10754, 10761 or 10767 that may be displayed during
boot-up or just before scanning. This fault mainly happened with machines that have
been upgraded from software CRE_1206X to CRE_1309 in the field.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
01-2011
Software Update
08/2007
 2292 (+ 2268)
MEDNET, GSO Library* (500 kB size)
Software CRE_1309 has to be installed before installing
the software patch CRE_1309A
DD+DIS304.07E, Service Bulletin No. 18, Sporadic
Errors 10754, 10761, or 107 67 with SW CRE_1309:
Install Patch CRE_1309A
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Repair and Service
Software Releases and Patches
(Intranet Link
Agfa Company Confidential
/
Extranet
Link)
Chapter 3.8 / 10