DD+DIS150.06E
2.2.3

Service Client Menu "SW-Versioning"

The "software versioning" menu allows checking if the software version currently
Purpose
installed on the control PC corresponds to the version installed on the Digitizer.
(1) Select "General Device Info - SW Versioning".
Usage
(2) Select "Get installed version".
Digitizer
IMPORTANT:
The installed versions on the PMI board, IP Handling board (HDL Board) and Slow
Scan Board (SSC Board) have always to be the same.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
Node
Installed Version Version on PC
CRE_1509/1096 CRE_1509/1096
PMI Board/FPGA
CRE_1509
HDL Board
CRE_1509
SSC Board
Get installed version
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
Tools and auxiliary Means
Status
ok
ok
CRE_1509
ok
CRE_1509
Figure 10
CR 30-X
Repair and Service
NX Workstation
(Control PC)
5175_chapter32_001.cdr
Chapter 3.2 / 14