DD+DIS150.06E
(2) When finished with re-mounting, continue with:
• Basic Protective Earth Test. See section 4.3.1.
• Electrical Check according to National Regulations. See section 4.3.2.
• Verification of the successful power supply replacement. See section 4.3.3.
4.3.1
Basic Protective Earth Test
(1)
Use a multimeter that is capable to
measure a resistance < 1 Ohm.
(2)
Put one probe at the ground connector of
the mains input.
(3)
Put the other probe at a few metallic parts,
the customer could touch.
The electrical resistance has to be ≤ 0.2
Ohm.
Refer to Figure 252.
(4)
Insert the power cord at the digitizer side;
leave it disconnected at the wall side.
(5)
Put one probe at the ground connector of
the mains connector.
(6)
Put the other probe at a few metallic parts,
the customer could touch.
The electrical resistance in this case has
to be ≤ 0.3 Ohm.
Refer to Figure 253.
IMPORTANT:
In case the electrical resistance of one measurement exceeds the limits (0.2 Ohm
without mains cable; 0.3 Ohm with mains cable) check the digitizer for loose grounding
connections!
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
200
Figure 252
Figure 253
Chapter 3.5 / 104
Agfa Company Confidential
517511ai.cdr
517509ah.cdr

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents