DD+DIS150.06E
(1) Insert cassette with unexposed IP in the digitizer.
Verification:
(2) At the workstation select examination type "System diagnosis – Flatfield".
(3) Select "ID" to identify the image: Scanning starts.
(4) Confirm that the image arrives at the workstation.
(5) Reject the image.
The cassette edge sensor is replaced.
Result:
2.3
Replacing the Cassette Fixation Steel Band
The replacement instructions for the Cassette Fixation Steel band, spare part number*
CM+9 5175 5830 0, are enclosed to the spare part.
* The last digit in the spare part number indicates the spare part revision at release of this document.
When ordering, the actual revision of the spare part is delivered.
2.4
Replacing the Cassette Fixation Unit
The replacement instructions for the Cassette Fixation Unit, spare part number*
CM+9 5175 9040 2, are enclosed to the spare part.
* The last digit in the spare part number indicates the spare part revision at release of this document.
When ordering, the actual revision of the spare part is delivered.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Chapter 3.5 / 23