Cleaning and Disinfection

All appropriate policies and procedures should be followed to avoid
contamination of the staff, patients and device. All existing universal
precautions should be taken to avoid that the digitizer comes into contact
with potential contaminations. Details about cleaning can be found in the
following pages.
For cleaning and disinfection instructions of the plates and cassettes, refer to
the CR 30-X/CR 30-Xm Plates and Cassettes User Manual.
To clean the exterior of the digitizer:
Switch off the digitizer.
1
Remove the power plug from the socket.
2
Switch off the UPS, if installed.
Wipe the exterior of the digitizer with a clean, soft, damp cloth.
3
Use a mild soap or detergent if required but never use ammonia-based cleaner.
Caution:
Note: Do not open the digitizer for cleaning. No components inside the
digitizer require cleaning by the user.
Plug the power plug into the socket.
4
Switch on the UPS, if installed.
2386G EN 20120110
Make sure no liquid gets in the digitizer.
CR 30-X/CR 30-X
M
23