Scope of Delivery

The Full Leg/Full Spine Stand "ADC Easylift" consists of the following
assemblies:
Framework with Sprocket Holder
Counterweight
Slide wagon
Upper front cover
Lower front cover
and the following single parts packed in small PE-bags:
Handle, brake cable and sleeve
Belt and belt connection to slide wagon
Adjustable feet and holders
Cassette holder mounting set
Slide wagon locking screws
and documents:
installation manual
user manual
assembly checklist (please store on a secure place as this document
may be needed for later reference in case of trouble with the
ADC Easylift)

Installation Planning

A safe installation of ADC Easylift requires two technicians

Installation Site

Upon defining the installation site please take into account:
The ADC Easylift must be installed in front of a wall impervious to x-rays
The floor covering must be solid enough to stand the weight of the ADC
Easylift (approx. 100 kg)
The customer has to decide early enough if the ADC Easylift should be
operated from the left or right hand side.
On the operator's side a minimum of 50 cm free space is needed to be
able to insert large format cassettes
Make sure that you have at least 2 hours access to the x-ray room on the
day of installation