WS-Manual-001, WS-Manual-T-001, WS-Manual-002, WS-Manual-T-002 | Remarks for HF-

Cables, transducers and accessories

Cables, transducers and accessories which were tested and found to comply
with the collateral standard IEC60601-1-2 (EMC):
from;
to
Transfer point table;
transfer point wall stand
control room (light push button);
table input terminal
control room (lamp red);
table input terminal
control room (lamp yellow);
table input terminal
control room (door contact);
table input terminal
control room (Com A);
table input terminal
control room (Com B);
table input terminal
control room (ground);
table input terminal
CAUTION:
Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and
result in improper operation.
type;
maximum length
10 x AWG21 (0.5 mm
;
20 m
2 x AWG21 (0.5 mm
15 m
2 x AWG18 (1.0 mm
15 m
2 x AWG18 (1.0 mm
15 m
2 x AWG18 (1.0 mm
15 m
9 pin sub D;
20 m
Standard RS-232 cable (9
pin sub D);
20 m
2
1 x AWG8 (10 mm
;
15 m
emission and immunity | 61
remark
unshielded
2
)
not delivered
2
);
with the sys-
tem
not delivered
2
);
with the sys-
tem
not delivered
2
);
with the sys-
tem
not delivered
2
);
with the sys-
tem
shielded
shielded
mandatory
)
3252C EN 20191206 1610