SCPI Programming Reference

TEST:ALL?

Runs an instrument self-test and returns a pass/fail indication. The
prehensive than the
*TST?
You must remove all input connections to the instrument before running the full self-test.
Parameter
(none)
Run a self-test:
TEST:ALL?
Typical Response: +0 (pass) or +1 (one or more tests failed)
If one or more tests fail, the query returns +1 and stores an error in the error queue. For a complete list-
l
ing of the error messages related to self-test failures, see
Following the test, the instrument returns to the instrument state that existed before the self-test.
l
208
self-test.
Typical Return
(none)
TEST:ALL?
self-test is more com-
Self-test Error
Messages.