Odi Loopback Test - Keysight X Series Installation Notes

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Option ST1 and ST2 – IQ Streaming

ODI Loopback Test

Use the following procedure to run the loopback test to verify that the newly installed streaming
option hardware is working properly.
1. Connect a USB keyboard and mouse to the instrument.
2. Change the instrument mode and measurement to IQ Analyzer mode and the Streaming
measurement by pressing Mode/Meas, IQ Analyzer, Streaming, OK.
3. Insert the Loopback Dongle that is included with this kit (M8121-60004) into the rear panel
ODI port with the alignment tab pointing up, making sure that it positively snaps into place.
It is very important that the dongle is correctly inserted all the way into the rear panel ODI port.
It must click into place such that it cannot be removed without sliding the spring loaded latch to
release it.
4. Using the USB mouse, select Start, Keysight Connection Expert. You should then see the
connection expert, as seen in
Figure 30
Keysight Connection Expert
30
Figure
30.
Installation Note N9032-90010

Advertisement