Preparation for Accurate Measurement (Executing Correction)
Executing Offset Correction
Executing Offset Correction
Turning ON/OFF the correction function
When you turn ON the offset correction, assuming that the measured value
before correction is
value is compensated as
To turn ON/OFF the offset correction, use the following command.
•
You cannot turn ON/OFF the primary parameter and secondary parameter
separately. However, if you set the correction value to 0, the state is actually
the same as OFF even if the offset correction is ON. Therefore, in practice you
can realize separate ON/OFF states by setting the correction value for either
parameter to 0.
Setting up the correction data
To set up the offset correction data, use the following command.
•
The entered value is set as the offset correction data for the measurement
frequency at the time of the entry.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Meas
[:SENSe]:CORRection:OFFSet[:STATe]
[:SENSe]:CORRection:OFFSet:DATA
and the offset correction data is
.
Meas Offset
–
on page 247
on page 246
, the measured
Offset
53