Troubleshooting the Emulation Probe
Problems with the Emulation Probe
If a performance verification test fails
There are some things you can do if a failure is found on one of the tests.
Details of the failure can be obtained through using a -v value ("verbose
level") of 9.
I f t h e p a r t i c u l a r f a i l u r e y o u s e e i s n o t l i s t e d b e l o w , c o n t a c t A g i l e n t T e c h n o l o g i e s f o r a s s i s t a n c e .
Test 1: Powerup PV Results
Failure of this test indicates a hardware problem with the emulation probe.
Contact Agilent Technologies for assistance.
Test 2: Emulation Module Port Feedback Test
Failure of this test indicates a hardware problem with the emulation probe.
Contact Agilent Technologies for assistance.
This test exercises the hardware which drives the connection to the emulation
module. It does not test the module/probe interconnect cable.
The test is not executed if the emulation probe is connected to an emulation
module.
Test 3: Run Control FPGA Test
Test 4: Run Control Clock Test
Failure of these tests indicates a hardware problem with the emulation probe.
Contact Agilent Technologies for assistance.
If the emulation probe fails one of these tests, it may have been damaged by
electrostatic discharge through the target cable. To prevent such damage in
the future, follow standard ESD preventive practices.
Test 5: Break In and Trigger Out SMB Feedback Test
Before returning to Agilent Technologies, check to ensure that you have
connected a good cable between the two SMB connectors.