RF Output Power Correction Constants (Test 47)

This procedure adjusts several correction constants that can improve the output power
level accuracy of the internal source. They are related to the power level, power slope,
power slope offset, and the ALC roll-off factors among others.
Analyzer warmup time: 30 minutes.
Required Equipment and Tools
Description
Power meter
Power sensor
Power splitter (2)
Attenuator 20 dB
GPIB cable
RF cable 24-inch
Antistatic wrist strap
Antistatic wrist strap cord
Static-control table mat and earth ground wire
Additional Equipment for Option 006 Analyzers
Power sensor
Procedure
1. If you just completed
start at step 8 in this procedure.
2. Press
Preset
Local
3. Press
SET ADDRESSES
Local
address is 13. Refer to the power meter manual as required to observe or change its
GPIB address.
4. Press
POWER MTR:438A/437
power meters. Choose the appropriate model number. (Use the 438A/437 selection if the
power meter is an E4419B or E4418B.)
NOTE
If you are using the 438A power meter, connect the 8482A power sensor to
channel A, and the 8481A power sensor to channel B.
Chapter 3
HP/Agilent Part or Model Number
436A/437B/438A or E4418B/4419B
8482A
11667A Option 001
8491A Option 020
10833A
11500B
9300-1367
9300-0980
9300-0797
8481A
"Sampler Magnitude and Phase Correction Constants (Test 53),"
SYSTEM CONTROLLER
ADDRESS: P MTR/GPIB
to toggle between the HP/Agilent 438A/437 and 436A
Adjustments and Correction Constants
RF Output Power Correction Constants (Test 47)
.
. The default power meter
3-11