8753ES Specifications and Characteristics
Instrument Specifications

Test Port Output

Table 1-10 Test Port Output
8753ES Test Port Output
Description
Frequency
Range
Standard
Option 006
Resolution
Stability
Standard
Option 1D5
CW Accuracy
a
Output Power
(above 300 kHz)
b
Level Accuracy
Standard
Option 075
Maximum Leveled Power
Standard
Option 075
Option 014
d
Power Range
Standard
Option 075
Option 014
Power Sweep Range
Standard
Option 075
Option 014
a. Source output performance on port 1 only. Port 2 output performance is a characteristic.
b. Absolute power accuracy at a given power level. Includes absolute accuracy and relative
flatness across frequency.
c. At any given frequency, the achievable power while remaining leveled. Applies to CW
mode only.
d. Power to which the source can be set and phase lock is assured.
1-12
Specification
30 kHz to 3.0 GHz
30 kHz to 6.0 GHz
1 Hz
±
10 ppm
±1.0 dB
±1.0 dB
c
−85 to +10 dBm
−85 to +8 dBm
−85 to +8 dBm
25 dB
23 dB
23 dB
Supplemental Information
10 kHz to 3 GHz, typ.
10 kHz to 6 GHz, typ.
±
7.5 ppm, 0˚ to 55˚C, typ.
±
3 ppm/year, typ.
0.05 ppm, 0˚ to 55˚C, typ.
±
0.5 ppm/year, typ.
±
at 25˚C
5 ˚C
at 0 dBm output level
at 0 dBm output level
+10 dBm, char.
+8 dBm, char.
+8 dBm, char.
33 dB, typ.
31 dB, typ.
31 dB, typ.