System:communicate:network:printer:net:graphoption - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example

:SYSTem:COMMunicate:NETwork:PRINter:NET:GRAPhoption

:SYSTem:COMMunicate:NETwork:PRINter:NET:GR
APhoption
This command is one of the commands to register the network printer setup. To
register a printer setup, the following commands must be entered:
:SYSTem:COMMunicate:NETwork:PRINter:NET:NAME
:SYSTem:COMMunicate:NETwork:PRINter:NET:IPADdress
:SYSTem:COMMunicate:NETwork:PRINter:NET:TEXToption
:SYSTem:COMMunicate:NETwork:PRINter:NET:GRAPhoption
:SYSTem:COMMunicate:NETwork:PRINter:NET:TYPE
:SYSTem:COMMunicate:NETwork:PRINter:SET
You can register maximum 4 setups.
This command sets the graphic output option of lpr command which is sent to the
remote printer via your print server.
If you do not enter the :SYST:COMM:NET:PRIN:SET command, *RST clears this
setting. Query returns the present setting.
:SYSTem:COMMunicate:NETwork:PRINter:NET:GRAPhoption option
Parameter
Type
option
string
option <newline><^END>
OUTPUT @Hp4155;":SYST:COMM:NET:PRIN:NET:GRAP '-h-l'"
OUTPUT @Hp4155;":SYST:COMM:NET:PRIN:NET:GRAP?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
graphic output option of lpr command.
SCPI Commands
5-309

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c

Table of Contents