Agilent Technologies 4155C Command Reference Manual page 417

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Error -105
Error -108
Error -109
Error -110
Error -111
Error -112
Error -113
Error -114
Error -120
Error -121
GET not allowed
A group execute trigger was received within a program message.
Parameter not allowed
Too many parameters for the command were received.
Missing parameter
Fewer parameters were received than required for the command.
Command header error
An error was detected in the header. This error message is reported if the
4155C/4156C cannot determine the more specific header errors -111 through -114.
Header separator error
An illegal character for a header separator was received; for example, no white
space between the header and parameter.
Program mnemonic too long
A keyword in the command header contains more than twelve characters.
Undefined header
An undefined command header was received; for example, *XYZ.
Header suffix out of range
The value of a numeric suffix attached to a program mnemonic is out of range; for
example, :PAGE:CHAN:SMU7:MODE V specifies illegal SMU number 7.
Numeric data error
Numeric (including the nondecimal numeric types) data error. This error message is
reported when the 4155C/4156C cannot determine the more specific errors -121
through -128.
Invalid character in number
An invalid character for the data type was received; for example, an alphacharacter
was received when the type was decimal numeric.
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Error Messages
Command Error
6-5

Hide quick links: