Page:measure:sampling:scondition:name - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Parameter
Query response
Example
Syntax
Parameter
Query response
Example
Parameter
Type
LOW
character
HIGH
character
ABSLow
character
ABSHigh
character
LOW | HIGH | ABSL | ABSH <newline><^END>
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:SCON:EVENT HIGH"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:SCON:EVENT?"
ENTER @Hp4155;A$

:PAGE:MEASure:SAMPling:SCONdition:NAME

This command selects the variable NAME to use with the THReshold and EVENt
to define the stop condition of sampling measurement.
:PAGE:MEASure:SAMPling:SCONdition:NAME var_name
Parameter
Type
var_name
string
var_name <newline><^END>
var_name is string response data, but does not contain double quote characters at the
beginning and end of the string.
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:SCON:NAME 'VD'"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:SCON:NAME?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure:SAMPling:SCONdition:NAME
Explanation
NAME value < THReshold
NAME value > THReshold
NAME | value | < | THReshold |
NAME | value | > | THReshold |
Explanation
data variable name
SCPI Commands
5-217

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c