:PAGE:DISPlay[:SETup]:DVARiables[:SELect]

This command selects the additional data variables for the graph or list. The
coordinate values of these variables will be displayed according to the position of
the marker.
If the specified variable name already exists, this command overwrites the old data
with the newly defined data.
A maximum of two data variables can be used. If this limit is exceeded, Cannot
define more than 2 displayed data vars. error (-184) is generated.
On the DISPLAY:DISPLAY SETUP page, this command fills the DATA
VARIABLE blank fields from upper field to lower.
At *RST, no variable is selected.
:PAGE:DISPlay [ :SETup] :DVARiables [ :SELect] var_name { ,var_name}
Parameter
Type
var_name
string
var_name { ,var_name} <newline><^END>
Response is a list of data variables separated by commas, and is string response data,
but does not contain double quote characters at the beginning and end of the string.
OUTPUT @Hp4155;":PAGE:DISP:DVAR 'VD','VG'"
OUTPUT @Hp4155;":PAGE:DISP:DVAR?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:DISPlay[:SETup]:DVARiables[:SELect]
Explanation
data variable name
SCPI Commands
5-123