Page:channels[:Cdefinition]:Pgu<N>:Standby - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example
Syntax
Parameter
Query response
Example
At *RST, this value is not defined.
:PAGE:CHANnels [ :CDEFinition] :PGU<n>:MODE V | VPULse
Parameter
Type
V
character
VPULse
character
V | VPUL | DIS <newline><^END>
If PGU is not used, the response data may be DIS(DISable).
OUTPUT @Hp4155;":PAGE:CHAN:PGU1:MODE VPUL"
OUTPUT @Hp4155;":PAGE:CHAN:PGU1:MODE?"
ENTER @Hp4155;A$
:PAGE:CHANnels[:CDEFinition]:PGU<n>:STANdby
This command sets the standby (STBY) function of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
At *RST, this value is OFF.
:PAGE:CHANnels [ :CDEFinition] :PGU<n>:STANdby OFF | ON | 0 | 1
Parameter
Type
OFF or 0
boolean
ON or 1
boolean
0 | 1 <newline><^END>
OUTPUT @Hp4155;":PAGE:CHAN:PGU1:STANDBY ON"
OUTPUT @Hp4155;":PAGE:CHAN:PGU1:STANDBY?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:CHANnels[:CDEFinition]:PGU<n>:STANdby
Explanation
voltage output mode
voltage pulse output mode
standby function is OFF
standby function is ON
SCPI Commands
Explanation
5-87