Syntax
Remarks
Example
Syntax
Example
See also
SCPI Commands

:PAGE:SCONtrol:KSWeep[:STARt]

:PAGE:SCONtrol:KSWeep[:STARt]
This command enables the start of the Knob SWeep operation.
After this command is executed, you start operation by rotating the front panel knob.
This command changes the present display page to the KNOB SWEEP page.
Executing this command is about the same as pressing Green front-panel key, then
pressing Single front-panel key.
This command does not have query form.
:PAGE:SCONtrol:KSWeep [ :STARt]
:PAGE:SCON:KSW command is not available when the 4155C/4156C screen is set
to the update disable state by :DISP OFF (or 0) command. To use this command,
enter :DISP ON (or 1) command.
OUTPUT @Hp4155;":PAGE:SCON:KSW"
:PAGE:SCONtrol[:MEASurement]:APPend
This command starts the APPend measurement operation.
The executed measurement mode, SWEep or SAMPling, is selected by the
:PAGE:CHANnels[:CDEFinition]:MODE command.
This command changes the present display page to the GRAPH/LIST: LIST or
GRAPH/LIST: GRAPH page, which is selected by the
:PAGE:DISPlay[:SETup]:MODE command.
Executing this command is about the same as pressing Append front-panel key.
This command does not have query form.
:PAGE:SCONtrol [ :MEASurement] :APPend
OUTPUT @Hp4155;":PAGE:SCON:APP"
:PAGE:CHANnels[:CDEFinition]:MODE and :PAGE:DISPlay[:SETup]:MODE