Page:scontrol:trigger:output:level - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Example
Syntax
Parameter
Example
Syntax
Example
SCPI Commands

:PAGE:SCONtrol:TRIGger:OUTPut:LEVel

:PAGE:SCONtrol:TRIGger:OUTPut:LEVel
This command outputs a level trigger from the external trigger output terminal.
This command does not have query form.
:PAGE:SCONtrol:TRIGger:OUTPut:LEVel HIGH | LOW
Parameter
Type
HIGH
character
LOW
character
OUTPUT @Hp4155;":PAGE:SCON:TRIG:OUTP:LEV HIGH"
:PAGE:SCONtrol:TRIGger:OUTPut[:PULSe]
This command outputs a pulse trigger from the external trigger output terminal.
This command does not have query form.
:PAGE:SCONtrol:TRIGger:OUTPut [ :PULSe] POSitive | NEGative
Parameter
Type
POSitive
character
NEGative
character
OUTPUT @Hp4155;":PAGE:SCON:TRIG:OUTP POS"
:PAGE:SCONtrol:ZERO
This command measures the zero offset that will be used for zero offset
cancellation.
Executing this command is about the same as pressing Green front-panel key, then
pressing Stop front-panel key.
This command does not have query form.
:PAGE:SCONtrol:ZERO
OUTPUT @Hp4155;":PAGE:SCON:ZERO"
5-246
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
output level is changed to HIGH
output level is changed to LOW
Explanation
output level is changed to LOW-HIGH-LOW
output level is changed to HIGH-LOW-HIGH

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c

Table of Contents